Abstract

The electrocatalytic activities of a series of compact graphites modified with microquantities of platinum metals (Pd or Pt+Pd) towards the electrochemical reduction of hydrogen peroxide were characterised. Operational parameters such as the optimal working potential, the influence of temperature and the resulting electrode characteristics were examined. The benefits of using graphite modified with Pt+Pd (mixture ratio 30%:70%) as the basic transducer in a glucose biosensor with improved sensitivity were demonstrated. It was proven that, under the working conditions chosen, the selected electrode (whether bare or covered with an enzyme layer) did not respond to any glutathione, uric acid or ascorbic acid (which all normally occur in biological fluids) present.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.