Abstract

Loudness growth functions characterize how the loudness percept changes with current level between the threshold and most comfortable loudness level in cochlear implant users. Even though loudness growth functions are highly listener-dependent, currently default settings are used in clinical devices. This study investigated whether electrically-evoked auditory steady-state response amplitude growth functions correspond to behaviorally measured loudness growth functions. Seven cochlear implant listeners participated in two behavioral loudness growth tasks and an EEG recording session. The 40-Hz sinusoidally amplitude-modulated pulse trains were presented to CI channels stimulating at a more apical and basal region of the cochlea, and were presented at different current levels encompassing the listeners' dynamic ranges. Behaviorally, loudness growth was measured using an Absolute Magnitude Estimation and a Graphic Rating Scale with loudness categories. A good correspondence was found between the response amplitude functions and the behavioral loudness growth functions. The results are encouraging for future advances in individual, more automatic, and objective fitting of cochlear implants.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.