Abstract
Functional graded materials (FGMs) are widely used for hip plant component materials because of their specific properties in a specific design to meet the requirements of the hip joint system. Stem is manufactured with the concept of making new FGM using three different composition of Ti, CrCo, and HA. This has led to the development of a suitable design for functional femoral parts that can be used for a long time. The new FGM design has helped to reduce the protection of the stress and the associated stress that exists on the interface. After changing in composition of FGM, the natural frequency and deformation response exhibited by our model were quite satisfactory. The deflection comes out to be 0.33 mm at the frequency of 15 Hz. The equivalent von Mises stress is found to be 19.5 MPa. These values are closely matching with previous studied. Fatigue study was carried based on “Goodman” theory. This gave promising results. The safety factor found is more than one, which indicates that the part is safe under design. The anticipated life of stem model is 0.3284 × 105 cycle.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.