Abstract

This paper demonstrates that a hand-powered egg beater can be modified to serve as a centrifuge for separating plasma from human whole blood. Immunoassays used to diagnose infectious diseases often require plasma from whole blood, and obtaining plasma typically requires electrically-powered centrifuges, which are not widely available in resource-limited settings. Human whole blood was loaded into polyethylene (PE) tubing, and the tubing was attached to the paddle of an egg beater. Spinning the paddle pelleted the blood cells to the distal end of the PE tubing; the plasma remained as the supernatant. A cholesterol assay (run on patterned paper) demonstrated the suitability of this plasma for use in diagnostic assays. The physics of the system was also analyzed as a guide for the selection of other rotating systems for use in centrifugation. Egg beaters, polyethylene tubing, and paper are readily available devices and supplies that can facilitate the use of point-of-care diagnostics at sites far from centralized laboratory facilities.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.