Abstract

One of the most essential difficulties in heavy-ion dosimetry by means of thermoluminescence dosemeters (TLDs)--often seen as a serious disadvantage of TLD utilisation--regards the changing TL-efficiency with increasing linear energy transfer (LET) of the particle. This behaviour leads to a falsification of absorbed dose that can be significant for many applications, e.g. in space or radiotherapeutic dosimetry. The high-temperature TL emission of LiF:Mg,Ti TL detectors can be exploited to obtain information about the LET of the heavy-ion radiation field under study. The high-temperature ratio (HTR) is used as a parameter to determine average LET. To correct the absorbed dose according to the TL-efficiency, the detailed dependence of HTR- and TL-efficiency on LET was recorded. These investigations were accomplished at the Heavy Ion Medical Accelerator (HIMAC) in Chiba, Japan, with a variety of high-energy ion beams (helium, carbon, neon, silicon and iron) ranging in LET from 2.2 to 393 keV microm(-1). The obtained relationships HTR vs. LET and TL-efficiency vs. LET were combined into a TL efficieny vs. HTR relationship. This enables correction of the absorbed dose (HTR-B method). The methodology is demonstrated by means of TLD 700 ((7)LiF:Mg,Ti) measurements in carbon beams of 290 and 400 MeV n(-1) available from HIMAC.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.