Abstract

Objective. In this study we examine the efficacy of anti-retraction devices in preventing microbial contamination of dental unit water lines (DUWLs). Methods. The study was performed on 54 randomly selected DUs in use at private and public institutions for over six months. The selected DUs were all currently commercially available. To evaluate the efficiency of anti-retraction devices, two different methods were employed, mechanical and microbiological. The ADA/ANSI specification #47 (corresponding to a water retraction of less than 40.3 μl) was used for the evaluation of the retraction determined using the mechanical method; the presence/absence of test microorganisms in DUWLs upstream the turbine after the test was used for evaluating the microbiological assay. Results. Both evaluation methods indicated correct amounts of retraction and prevention of DUWL contamination occurred in only two cases (3.7%). Correct retraction was noted in six other dental units (DUs), but their DUWLs became contaminated during testing. Six other produced the opposite results—improper retraction, yet no contamination. The remaining 40 DUs (74.0%) failed both evaluations. Conclusion. The results showed that the overwhelming majority of the anti-retraction devices did not prevent retraction when the turbine stopped running, leading to a contamination of the water lines, and to a consequent possible cross-contamination of the patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.