Abstract

Introduction: The concurrent literature reflects wide variations in the use of response filter for acquiring ocular vestibular-evoked myogenic potentials (oVEMPs). However, there is dearth of published reports on the effects of changes in response filter on oVEMP. Therefore, the present study aimed at evaluating the effect of variations in the response filter on oVEMP and identifying the optimum filter set for its clinical recording. Methods: Contralateral oVEMPs were elicited in response to 500 Hz tone bursts from thirty healthy individuals. The low-pass filters used were 500, 700, 1000, 1500, 2000, and 3000 Hz, and the high-pass filters used were 1, 10, and 30 Hz, in all possible combinations. Results: There was a significant reduction in n1- and p1-latencies with increase in high-pass and low-pass filters (P

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.