Abstract

Quantitative data on the effect of the electrode geometry on the signal strength and the signal-to-noise ratio is given. The measurements are affected by the unavoidable presence of stray capacitance. Best results are achieved for short and narrow electrodes arranged in an antiparallel configuration and separated by a minimal gap, which determines the dimensions of the actual detection volume. Limits of detection between 150 and 250 microg l(-1) and separation efficiencies from 13,000 to 17,000 theoretical plates were achieved for six inorganic cations (NH(4)(+), K(+), Ca(2+), Na(+), Mg(2+)and Li(+)) with electrodes of 1 mm width and a detection gap of 0.5 mm (separation channel length: 7.5 cm) when operating the detector at 20 V(pp) and 500 kHz. The analyses of all major inorganic cations in tap and rain water samples were demonstrated for the first time in microchip electrophoresis with contactless conductivity detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.