Abstract

Effects of sodium fluoride (NaF) mouth rinse solutions on torsional properties of a miniscrew implant were investigated. As-received Ti-6Al-4V miniscrew implants (AbsoAnchor [Dentos, Inc., Daigu, Korea]) were immersed in 0.1% or 0.2% NaF mouth rinse solution (pH 5.12 and 5.14, respectively) for 1 hour or 24 hours. Miniscrew implants selected as controls were not immersed. Each implant was subjected to increasing torque until fracture (n = 5 in sample groups). Mean moment and twist angle for fracture were compared using 1-way analysis of variance (ANOVA). Surfaces of implants after immersion were observed with a scanning electron microscope (SEM). Electron microprobe and micro-x-ray diffraction analyses were performed to obtain composition information about deposits on implant surfaces. Pits and cracks formed on the implant surfaces after immersion in both NaF mouth rinse solutions. Corrosion products, probably sodium aluminum fluoride (Na(3)AlF(6)), were observed on the implants after immersion in both NaF solutions for both time periods. There were no significant differences for mean torque (P = 0.063) and twist angle (P = 0.696) at fracture compared with control implants. Although titanium alloy miniscrew implants corroded slightly from immersion in 0.1% or 0.2% NaF solutions, mouth rinsing by patients with the same fluoride solution concentrations should not cause deterioration of their torsional performance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.