Abstract

In studies to date, the effectiveness of the directional microphone has been investigated independently of the signal processing scheme used in the hearing aid. In addition, the number and placement of the background noise speakers can create an advantage for a particular polar pattern (i.e., cardioid, supercardioid, and hypercardioid) in any laboratory design. With these considerations in mind, the purpose of this investigation was twofold: (1) to determine the effect of different amplitude-compression release times on speech perception ability in noise, measured with directional microphone hearing aids, and (2) to determine the impact of environment (classroom vs anechoic chamber) on those measures. Ten subjects with mild to moderate sensorineural hearing loss participated. Using an eight-speaker complex sound field, speech perception was assessed in an anechoic chamber and a typical classroom environment. None of the release times resulted in superior performance in either the anechoic or classroom environment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.