Abstract

Several anomalies, e.g., in peak shape, migration time, and baseline drift, all due to pressure-driven backflow, were previously reported to occur during serial injection on capillary electrophoresis (CE) chips. Since these anomalies were worse for polydimethylsiloxane (PDMS) microchips than for glass microchips, reproducible data on PDMS microchips were difficult to obtain. In this paper, we found that these problems were affected by the hydrophilic or hydrophobic properties of the reservoirs on the microchip and demonstrated that these anomalies were reduced by converting the hydrophobic properties of the reservoirs on the PDMS microchip into hydrophilic ones. Thus, compared with hydrophobic reservoirs, hydrophilic reservoirs were suitable for the formation of a stable plug. Several chip designs were suggested to reduce these pressure-driven backflows.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.