Abstract

The acrylic scintillator which was used for several years as the active medium in the Axial Field Spectrometer uranium calorimeter at the CERN Intersecting Storage Rings was found to be severely damaged. The observed light output has decreased on the average by a factor of 2, and the attenuation length by a factor of 3, causing a severe degradation of the energy resolution of the system for the detection of both electromagnetic and hadronic showers. It could be shown that the effects are due to the radiation induced by uranium decay. The fact that the integrated dose obtained from this process (50 Gy) is 2 orders of magnitude smaller than what was currently believed to be a safe level, strongly suggests that the dose rate is a dominating factor in these radiation damage problems. It could be shown that the radiation resistivity is improved by at least an order of magnitude if contact with oxygen is avoided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.