The performance of the aerodynamic particle sizer (APS) in an atmosphere that differs from standard calibration conditions was studied. The response of the APS for monodisperse polystyrene latex and dioctylphthalate particles ranging from 2.4 to 15.1 μm in diameter was measured in air, argon, and nitrous oxide atmospheres. The measurements indicated that particles in argon and in nitrous oxide accelerate faster than those in air. In order to postulate the mechanisms and interpret the observed results, the effects of both gas viscosity and gas density on instrument performance were considered. The particle density correction factor previously given by Wang and John (1987) (Aerosol Science and Technology. 6:191–198) was extended to include gas property effects. Good agreement was obtained between the new factor and the experimental data. Existing data obtained at a low pressure in air were also compared successfully with the developed model. Expected applications of the work are the use of the APS in an arb...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.