Abstract

The effect of strain upon the fluorescence intensity ratio technique of optical fiber based temperature sensing has been investigated using a series of different Yb3+- and Nd3+-doped optical fibers. Measurements were conducted with each of the doped fibers for temperatures between approximately 10 and 140 °C and for strain values between 50 and 2100 με. Results suggest that the strain-temperature cross sensitivity for the fluorescence intensity ratio method is minimal, with an applied strain of 1000 με inducing temperature errors of less than 0.4 °C.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.