Abstract
Asymmetrical flow field-flow fractionation is a rising useful technique to separate and characterize macromolecules. Elution behaviors of allergen protein ovalbumin and immunoglobulin E were studied here. Effects of flow rates, kinds of carrier solutions, and injection amounts were investigated in the desire to experiment and verify the effect of every factor and to validate asymmetrical flow field-flow fractionation theory. Results suggested that cross-flow is the most important factor among all the factors that have been studied. Too-low focus flow (0.5 ml/min) and outflow (0.4 ml/min) are harmful for protein separation and characterization. Smaller size ovalbumin (with a monomer is 11.4 nm) is more prone to be influenced compared with the bigger analyte immunoglobulin E (with a monomer diameter is 32.1 nm) by flow rates. The adopted types of carrier fluids did not change the retention behavior of the two proteins. Still, protein aggregation and membrane adsorption should be paid more attention to under different carrier fluids.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.