Abstract
This study tested the hypothesis that modular head seating in situ reduces the cement-stem interfacial strength. Femoral prostheses were implanted in eight pairs of cadaveric femurs; one femur of each pair received a stem for which the modular head was seated by dropping a weight from a given height (treated); the contralateral femur received a nonseated stem (control). Transverse sections were cut from four standardized locations on each implanted femur, and the yield shear stress and ultimate shear stress of the interface were determined from push-out tests. There was no significant difference in cement-stem interfacial strength between seated and nonseated prostheses. These results suggest that seating modular heads in situ has no deleterious effect on the acute interfacial strength of cemented femoral implants.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.