Abstract

The current study provides insights into the challenges of safely operating a magnetic resonance imaging (MRI)-guided focused ultrasound (MRgFUS) robotic system in a high-field MRI scanner in terms of robotic motion accuracy. Grid sonications were carried out in phantoms and excised porcine tissue in a 3T MRI scanner using an existing MRgFUS robotic system. Fast low-angle shot-based magnetic resonance thermometry was employed for the intraprocedural monitoring of thermal distribution. Strong shifting of the heated spots from the intended points was observed owing to electromagnetic interference (EMI)-induced malfunctions in system's operation. Increasing the slice thickness of the thermometry sequence to at least 8 mm was proven an efficient method for preserving the robotic motion accuracy. These findings raise awareness about EMI effects on the motion accuracy of MRgFUS robotic devices and how they can be mitigated by employing suitable thermometry parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.