Abstract

This study was designed to analyze flow pattern, velocity, and strain on the aortic wall of a glass aortic model during extracorporeal circulation, and to elucidate the characteristics of flow pattern in four aortic cannulas. Different patterns of large vortices and helical flow were made by each cannula. The high-velocity flow (0.6 m/s) was observed in end-hole cannula, causing high strain rate tensor (0.3~0.4 without unit) on the aortic arch. In dispersion cannula, a decreased strain rate tensor (less than 0.1) was found on the outer curvature of the aortic arch. In Soft-flow cannula (3M Cardiovascular, Ann Arbor, MI, USA), further decreased flow velocity (0.2 m/s) and strain (less than 0.2) were observed. In Select 3D cannula (Medtronic, Inc., Minneapolis, MN, USA), a high strain (0.4~0.5) was observed along the inner curvature of the aortic arch. In conclusion, end-hole cannula should not be used in atherosclerotic aorta. Particular attention should be paid both for selection of cannulas and cannulation site based on this result.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.