Abstract

We report a case of arcus senilis affecting the accuracy of intraocular lens (IOL) power calculation using intraoperative aberrometry. There was a large difference in the predicted IOL power (1.50 diopters) between the IOL power calculation using preoperative optical biometry and intraoperative IOL power calculation using intraoperative aberrometry (Optiwave Refractive Analysis). After the arcus senilis was adjusted for by manually measuring the white-to-white distance intraoperatively and that new value was inputted into the intraoperative aberrometer, the measurement matched the preoperative measurements and confirmed the correct choice of IOL. To our knowledge, this is the first reported case of arcus senilis affecting intraoperative IOL power calculation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.