Abstract

Evaluation of the Growth Stent-a stent consisting of two stent halves connected by reabsorbable sutures-for the treatment of aortic coarctation in infants. Surveillance study of 13 Growth Stents implanted in 12 patients aged 1-15 months (median 5 months). Body weight ranged from 3.4-12.8 kg (median 5.4 kg). Eight patients suffered from aortic (re-)coarctation, four of them from stenosis of the aortic anastomosis after a Norwood I procedure. The follow-up period was 24 months (11-51 months). Pressure gradients immediately after stent implantation decreased from 30 mm Hg (range 20-50 mm Hg) to 8 mm Hg (range 0-15 mm Hg). Five patients had one (3 pts) or two (2 pts) balloon dilations 3-28 months (median 12 months) after Growth Stent implantation. The median pressure gradient decreased from 25 mm Hg (range 15-30 mm Hg) to 15 mm Hg (range 5-25 mm Hg). Six patients received a large stent after 19-34 months. Median body weight was 11.8 kg (9.4-15 kg). The Growth Stent is suitable for the acute treatment of aortic coarctation in infants and can be overstented later on--if necessary--with a larger stent without causing restriction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.