Abstract

Mammography and ultrasound indicated a cancer of the right breast in a 77-year-old woman with a dual-chamber demand pacemaker. The patient was not pacemaker-dependent. She underwent breast 1.5T magnetic resonance imaging (MRI) (dynamic gradient echo sequence with Gd-DOTA 0.1 mmol/kg). Before the patient entered the MR room, the configuration of the device was changed (the response to magnet was switched from asynchronous to off and the rate-responsive algorithm was disabled). No relevant modifications of heart rhythm or rate were observed during the MR examination. No symptom was reported. Immediately after the examination, the pacemaker interrogation showed neither program changes nor alert warnings. MRI detected a bifocal cancer in the right breast which allowed tailored breast-conserving treatment to be initiated. Histopathology confirmed a bifocal invasive ductal carcinoma.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.