Abstract

The need for repositioning of surface coils and patients in MR examinations of the cervical and thoracic spine prolongs examination time. A new receiver design is proposed which overcomes this problem. The device is composed of two actively decoupled receiver coils mounted on the frame of a Philadelphia collar. These coils may be used separately to image either the thoracic or cervical spine or together to produce larger field-of-view images of the combined region. Signal-to-noise ratios of the separate cervical and thoracic spine images are not degraded as a result of mounting the receivers together. The full cervical and thoracic region is shown to be imaged at a signal-to-noise ratio significantly higher than that afforded by the body coil. A retrospective review of our case load suggests that a time saving could be achieved in approximately 1/3 of spine examinations by using this coil.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.