Abstract
Drift chambers have been installed for luminosity measurements in intersection 5 of the SPS accelerator working in p p colliding mode. The required electronics is described. The system is able to process up to 16 hits per wire with a double pulse resolution of 40 ns; drift time and current division, with 1.25 ns and 1.6% resolution respectively, are recorded. Transconductance preamplifiers and discriminators are directly mounted on the chamber; 160 m of twisted-pair cable bring the signals to the digitizer unit. Coarse time is measured using RAM techniques, while fine time is obtained by means of a microstrip delay associated with a 100 K ECL priority encoder. Current division used a single 50 MHz Flash ADC which allows 26 dB dynamic range with 6 bit resolution. First operational results are reported.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.