Abstract

The screening of irregular red-cell antibodies on 20.200 recipient samples was carried out simultaneously on continuous flux AUTO-ANALYZER (LIP-bromeline) and GROUPAMATIC (TPC-bromeline). From results it appears: --an equivalent sensibility of both technologies to detect Rh, Kell, Duffy and Kidd antibodies; --a higher sensibility of Groupamatic to detect particularly Lewis and P1 antibodies; --a clear superiority of TPC compared to bromeline technique which is commonly used with Groupamatic. Small modifications of the standard TPC system allow to associate, in routine work, detection of irregular red-cell antibodies with other analyses, such as ABO Rh grouping and phenotyping. From this last point of view, a restricted test of coagglutination phenotyping has revealed to be very promising.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.