Abstract
Lithium fluoride thermoluminescent (TL) detectors, with Li-7 isotope and various activators (MTS-7 LiF:Mg,Ti, MTT-7 LiF:Mg,Ti with enhanced Ti concentration and MCP-7 LiF:Mg,Cu,P) were used for dosimetry of heavy charged particles, within the ICHIBAN experiment. The microdosimetric model has been applied to calculate detection efficiency, eta, relative to gamma-ray dose, of these detectors after proton and heavy charged particle (HCP) irradiation for ion charges ranging from Z = 1 to Z = 6 and in the energy range from 0.3 to 20 MeV amu(-1). The calculated ratio eta(MCP-7)/eta(MTS-7) lies in the range between 0.2 and 1.0 for protons and between 0.2 and 0.4 for HCP with Z > 1. The calculated value of eta(MTT-7)/eta(MTS-7) for protons was found range between 1.0 and 1.45 and, for Z > 1, between 1.3 and 2. These relationships can be applied to derive information about the 'effective LET' in an unknown HCP field and to correct the TLD readings for dose evaluation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.