Abstract
For single plane implants with gold seeds, assuming a matrix distribution for the seeds, the optimum separation of the matrix elements and the optimum ratio of the activity of the seeds in the area to that of those in the periphery were calculated, so that the uniformity of dose distribution is within plus or minus 10%. Area dosages were also computed and are tabulated. The dose variations in double plane implants and volume implants also were studied and the results are presented. Dosage graphs are given for both cases. The theoretical dose distributions were verified for a few implants experimentally using LiF thermoluminescence dosimeters. (auth)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.