Abstract

Strong fields of thermal and epithermal neutrons as used for Boron Neutron Capture Therapy (BNCT) show a relatively small gamma dose. However, for accurate dose estimation, both the neutron as well as the gamma component has to be considered. Different approaches have been studied with various thermoluminescence detectors (TLDs). As a result, CaF 2:Tm detectors (TLD-300) are attributed to be most insensitive to thermal neutrons but not tissue-equivalent. In this paper we focus on the measurement of the gamma dose with TLD-300 in mixed irradiation fields. To overcome the disadvantage of being not tissue-equivalent, a study on energy compensation filters was performed. The paper comprises also calibrations of TLD-300 using energy compensation filters at the PTB-reference field for thermal neutrons, and first test experiments at the TRIGA Mainz including simulations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.