Abstract

This study was aimed to determine the dosimetry characteristics of the EBT3 film on a small field electron beam. The experiments were conducted using Linac Synergy-Platform with 6, 8, 10, 12, and 15 MeV. EBT3 films were cut into 3.2×2.5 cm2 in size and were irradiated with varied dose of 0 to 500 cGy on 1×1, 2×2, 3×3, 5×5, 8×8 and 10×10 cm2 radiation field sizes. The irradiated film was scanned using Epson Perfection V700 flatbed scanner to evaluate the pixel value. Then, the calibration curve was generated from netOD and given a dose. The uncertainty and sensitivity of the EBT3 film were evaluated using in-house and Image-J software to check the consistency of the result. The results showed that the largest uncertainty of characteristics curve was 4.6% at 1×1 cm2. Whereas the sensitivity value tends to be identical for radiation dose greater than 100 cGy for all energy, it depends on energy for radiation dose less than 100 cGy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.