Abstract

The detail signal-to-noise ratio model of radiographic imaging is quantitatively analyzed in terms of its accuracy in describing observer threshold perceptibility of radiographic detail. The model is found to adequately describe the effects of magnification, scatter radiation, and system resolution on observer threshold perceptibility. However, it is shown that the model does not apply in screen/film radiography for very low contrasts and high scatter conditions due to insufficient optical density contrast. The dose-to-information conversion efficiency of a radiographic imaging system is defined and the effects of magnification, scatter, resolution, image processing, detector efficiency, grids, patient table support, field size, and geometry on the dose efficiency of the imaging system are investigated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.