Abstract
Dose and transmission factor values (TF) were measured for four typical radiotherapy treatments in which block protections are employed, using ionization chambers and thermoluminescent dosimetry. Dose calculations using Clarkson method and the hospital planning system were compared to the measurements. Transmission factor and dose mean values measured with ionization chamber and thermoluminescent dosimeters were statistically consistent, and differences between Clarkson method and treatment planning dose calculations showed up to 5% disagree. When experimental and theoretical doses are compared, differences up to 35% have been found.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.