Abstract

Traveling wave ion mobility (TWIM) mass spectrometry (MS) is a powerful method for the structural and conformational analysis of proteins and peptides, enabling the differentiation of isomeric peptides (or proteins) that have the same sequence but are modified at different residues. In this study, the TWIM-MS technique was used to separate isomeric AuI metallopeptide ions that were formed by ZnII displacement from the parent zinc fingers (ZFs). The synthetic gold finger peptides were derived from the C-terminus of the HIV nucleocapsid p7 protein (NCp7-F2) and finger 3 of the Sp1 transcription factor (Sp1-F3). TWIM-MS enabled the acquisition of distinct product ion spectra for each isomer, clearly indicating the binding sites for the major conformers in the presence of multiple coordination possibilities. Collision cross-section measurements showed that the aurated peptide has a slightly more compact structure than the parent zinc compound NCp7-F2, which showed only one conformation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.