Abstract

Near-infrared (NIR) diffusion correlation spectroscopy (DCS) provides a novel tool for clinical monitoring of blood flow in deep tissues. While DCS has the advantages of non-invasiveness, high sensitivity and large penetration depth, accelerating its data acquisition process to realize real-time imaging is still an on-going task. We have proposed herein a 10-detection-channel DCS system for dynamic topography of the blood flow index. In the design, an imaging speed up to 1 frame/s can be achieved by taking the advantage of the concurrency of a multi-channel multi-delay designed hardware correlator, the improved NIR-sensitivity of a single photon avalanche photodiode detector and the high efficiency of the prior-guided fitting procedure for the blood flow index extraction. We firstly evaluated the flow-rate resolution, measurement fidelity and imaging depth of the system using phantom experiments, and then performed in-vivo experiments on human arm to demonstrate the ability of the proposed methodology to capture blood flow changes in terms of their shapes and positions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.