Abstract

Peroxynitrite, as a derivative of nitric oxide, is a potent oxidant. It reacts with several biological molecules, makes cellular and tissue damages, and is related with many diseases; therefore, it is of major concern in current medical research works. In this work, a special perm-selective cellulose acetate membrane sampler is used to implement flow injection analysis (FIA)/chemiluminscence (CL)-detection method for the detection of peroxynitrite with Luminol CL-reagent. Optimum detection conditions were established, and the permeability of peroxynitrite through cellulose acetate (CA) membrane, as well as the interference from matrix constituents were studied. The proposed method has the high sensitivity of the CL-detection and the selectivity of perm-selective membrane sampler. The obtained detection limit of 1×10 −11 M (without dialysis membrane) and 1×10 −10 M (with dialysis membrane), makes it possible to monitor the elusive peroxynitrite in biological samples. The mechanism of luminol CL-emission generated during oxidation by peroxynitrite and the kinetics of peroxynitrite decomposition were also studied using FIA/CL-detection set-up.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.