Abstract

We evaluated the diagnostic performance of the EMIT-tox serum benzodiazepine assay adapted to a Hitachi 717 analyzer (EMIT), the Abbott ADx serum benzodiazepine fluorescence polarization immunoassay (FPIA), and a radioreceptor assay (RRA) in 113 patients with suspected acute poisoning. The reference method was high-performance liquid chromatography with ultraviolet detection after solid-phase extraction. For the discrimination between negative and positive samples, the areas under the receiver-operating characteristic (ROC) curves were 0.976, 0.991, and 0.991 for EMIT (cutoff, 50-ng/mL diazepam), FPIA (cutoff, 12-ng/mL nordiazepam), and RRA (cutoff, 50-ng/mL diazepam), respectively. For the discrimination between non-toxic and toxic concentrations, the areas under the ROC curves were 0.896, 0.893, and 0.933, respectively. EMIT (with the cutoff lowered to 50 ng/mL), FPIA, and RRA can be reliably used to screen for the presence of benzodiazepines in serum, but in many cases they cannot discriminate between toxic and nontoxic concentrations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.