Abstract
Endobronchial valve (EBV) therapy is optimized in patients who demonstrate little or no collateral ventilation (CV). The accuracy of the Chartis System and visual assessment of high-resolution computerized tomography (HRCT) fissure completeness by a core radiology laboratory for classifying CV status was compared by evaluating the relationship of each method with target lobe volume reduction (TLVR) after EBV placement. Retrospective HRCT fissure analysis of a study population who underwent catheter-based measurement of CV followed by complete occlusion of the targeted lobe by EBV. Accuracy, sensitivity, specificity, positive predictive value and negative predictive value of the HRCT fissure analysis and the catheter-based measurement of CV for predicting TLVR was determined. Accuracy for correctly classifying TLVR with EBV was similar for Chartis System and HRCT fissure analysis (74 vs 77%). The sensitivity and specificity of the Chartis measurement were 86% and 61% and those of HRCT fissure analysis 75% and 79%. Patients with TLVR ≥ 350 mL had statistically significant improvement in respiratory function, exercise performance and quality of life measures. When evaluating patients for likelihood of successful EBV therapy, the Chartis System CV assessment and HRCT fissure analysis appear to have comparable accuracy. Both techniques were found to be beneficial for EBV procedure planning.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.