Abstract

The design and prototype tests of the Very Low Angle Detector (VLAD) for epithermal neutron scattering at low momentum transfers are presented. The detector will be installed on the VESUVIO spectrometer at the ISIS spallation neutron source for so-called High-energy Inelastic Neutron Scattering (HINS) experiments. It will consist of an array of Resonant Detectors (RD) covering the scattering range 1/spl deg/-5/spl deg/. The results of prototype tests with a reduced set of RD units demonstrate the feasibility of the measurement in the adverse environmental conditions of ISIS. Simulations of particle interactions using the GEANT4 code show that the envisaged detector design can meet the requirements in terms of efficiency, cross-talk between detection channels and neutron scattering events in the detector.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.