Abstract

Nimbus, Inc. and the University of Pittsburgh's School of Medicine have been collaborators developing rotary blood pump technology since 1992. Currently, a major focus is on an implantable left ventricular assist system (LVAS) that utilizes an electric powered axial flow blood pump. In addition to the blood pump, a major development item is the electronic controller and the control algorithm for modulating the pump speed in response to varying physiologic demands. Methods being used in developing the axial flow LVAS include the use of computational fluid dynamic modeling of the interior flow field of the pump, flow visualization of the flow field using laser based imaging, and computer simulation of blood pump-physiological interactions as well as an extensive in vivo test program. Results to date include successful in vivo tests of blood pumps with nonlubricated bearings and demonstrations of auto speed control using electrical current as the observable parameter.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.