Abstract

The development of an electron cyclotron resonance ion source for the heavy ion medical accelerator in Chiba (HIMAC) injector is reported. The HIMAC is a heavy ion medical accelerator for cancer therapy. The electron cyclotron resonance (ECR) ion source is expected to provide a long lifetime, easy operation, and easy maintenance for medical use. The NIRS-ECR ion source has a single closed ECR stage, and a microwave frequency of 10 GHz is applied. Under the present performance, the output electrical currents of the ions are 2500 eμA for He1+, 300 eμA for C2+, 480 eμA for Ne3+, and 110 eμA for Ar6+. Stability of the intensity is better than 2%. The transmission efficiency through a low-energy beam-transport line with an acceptance of 200 πmm mrad is more than 70%; the typical 50% and 90% emittances of the injection beam with 8 keV/u are 20 and 80 πmm mrad, respectively. These performances satisfy the requirements for radiotherapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.