Development of the 4D Phantom for patient-specific, end-to-end radiation therapy QA

  • Abstract
  • Literature Map
  • Similar Papers
Abstract
Translate article icon Translate Article Star icon

In many patients respiratory motion causes motion artifacts in CT images, thereby inhibiting precise treatment planning and lowering the ability to target radiation to tumors. The 4D Phantom, which includes a 3D stage and a 1D stage that each are capable of arbitrary motion and timing, was developed to serve as an end-to-end radiation therapy QA device that could be used throughout CT imaging, radiation therapy treatment planning, and radiation therapy delivery. The dynamic accuracy of the system was measured with a camera system. The positional error was found to be equally likely to occur in the positive and negative directions for each axis, and the stage was within 0.1 mm of the desired position 85% of the time. In an experiment designed to use the 4D Phantom's encoders to measure trial-to-trial precision of the system, the 4D Phantom reproduced the motion during variable bag ventilation of a transponder that had been bronchoscopically implanted in a canine lung. In this case, the encoder readout indicated that the stage was within 10 microns of the sent position 94% of the time and that the RMS error was 7 microns. Motion artifacts were clearly visible in 3D and respiratory-correlated (4D) CT scans of phantoms reproducing tissue motion. In 4D CT scans, apparent volume was found to be directly correlated to instantaneous velocity. The system is capable of reproducing individual patient-specific tissue trajectories with a high degree of accuracy and precision and will be useful for end-to-end radiation therapy QA.

Similar Papers
  • Research Article
  • 10.1118/1.2241491
TU‐B‐224C‐01: 4D Scanning
  • Jun 1, 2006
  • Medical Physics
  • S Jiang

The purpose of CT simulation in radiotherapy is to acquire patient geometrical information and to build a patient geometrical model for treatment planning. Errors in patient model caused by motion artifacts will influence all treatment fractions and therefore should be handled carefully. Due to the tumor respiratory motion, the captured tumor position and shape can be heavily distorted. The distortions along the axis of motion could result in either a lengthening or shortening of the target. The center of the imaged target can be displaced by as much as the amplitude of the motion.A newly developed technique that can reduce motion artifacts and provide patient geometry throughout the whole breathing cycle is called respiration‐correlated or 4D CT scan. The basic idea for 4D CT scan is that, at every position of interest along patient's long axis, images are over‐sampled and each image is tagged with breathing phase information. After the scan is done, images are sorted based on the corresponding breathing phase signals. Thus, many 3D CT sets are obtained, each corresponding to a particular breathing phase, and together constitutes a 4D CT set that covers that the whole breathing cycle. 4D CT scan has been developed at various institutions with slightly different flavors. In this lecture, we will provide an overview of various implementations of 4D CT scan.4D CT scan can be used to account for respiratory motion to generate images with less distortion than 3D CT scan. 4D images also contain respiratory motion information of tumor and organs that is not available in a 3D CT image. This technology can be used for respiratory‐gated treatment to identify the patient‐specific phase of minimum tumor motion, determine residual tumor motion within the gate interval, and compare treatment plans at different phases. It can also be used for non‐gated treatment planning to define ITV by combining gross tumor volume at all breathing phases or using a method called maximum intensity projection. Of course 4D CT will also play a vital role in the futuristic 4D radiotherapy where the tumor is tracked dynamically during the treatment using multi‐leaf collimator.Existing problems for 4D CT scan include the increased imaging dose, CT tube heating, and data management. More importantly, one has to keep in mind that 4D CT scan is not really 4D. Temporal information is mapped into one breathing cycle. Irregular respiration will cause artifacts in 4D CT images. Patient coaching can improve the regularity of breathing pattern and thus reduce the residual artifacts. However this issue still deserves further studies.Educational Objectives:1. Understand the origin and magnitude of motion artifacts in free breathing helical CT scan.2. Understand how 4D CT scan works.3. Understand how 4D CT can be used in radiotherapy.4. Understand the remaining artifacts in 4D CT scan and possible future improvements.

  • Research Article
  • 10.1118/1.2761475
WE-SAMS-AUD-01: 4D Scanning: Imaging/Planning
  • Jun 1, 2007
  • Medical Physics
  • Sb Jiang

The purpose of CT simulation in radiotherapy is to acquire patient geometrical information and to build a patient geometrical model for treatment planning. Errors in patient model caused by motion artifacts will influence all treatment fractions and therefore should be handled carefully. Due to the tumor respiratory motion, the captured tumor position and shape can be heavily distorted. The distortions along the axis of motion could result in either a lengthening or shortening of the target. The center of the imaged target can be displaced by as much as the amplitude of the motion. A newly developed technique that can reduce motion artifacts and provide patient geometry throughout the whole breathing cycle is called respiration‐correlated or 4D CT scan. The basic idea for 4D CT scan is that, at every position of interest along patient's long axis, images are over‐sampled and each image is tagged with breathing phase information. After the scan is done, images are sorted based on the corresponding breathing phase signals. Thus, many 3D CT sets are obtained, each corresponding to a particular breathing phase, and together constitutes a 4D CT set that covers that the whole breathing cycle. 4D CT scan has been developed at various institutions with slightly different flavors. In this lecture, we will provide an overview of various implementations of 4D CT scan. 4D CT scan can be used to account for respiratory motion to generate images with less distortion than 3D CT scan. 4D images also contain respiratory motion information of tumor and organs that is not available in a 3D CTimage. This technology can be used for respiratory‐gated treatment to identify the patient‐specific phase of minimum tumor motion, determine residual tumor motion within the gate interval, and compare treatment plans at different phases. It can also be used for non‐gated treatment planning to define ITV by combining gross tumor volume at all breathing phases or using a method called maximum intensity projection. Of course 4D CT will also play a vital role in the futuristic 4D radiotherapy where the tumor is tracked dynamically during the treatment using multi‐leaf collimator. Existing problems for 4D CT scan include the increased imagingdose,CT tube heating, and data management. More importantly, one has to keep in mind that 4D CT scan is not really 4D. Temporal information is mapped into one breathing cycle. Irregular respiration will cause artifacts in 4D CTimages. Patient coaching can improve the regularity of breathing pattern and thus reduce the residual artifacts. However this issue still deserves further studies. Educational Objectives: 1. Understand the origin and magnitude of motion artifacts in free breathing helical CT scan. 2. Understand how 4D CT scan works. 3. Understand how 4D CT can be used in radiotherapy. 4. Understand the remaining artifacts in 4D CT scan and possible future improvements.

  • Research Article
  • Cite Count Icon 146
  • 10.1118/1.2966347
Reconstruction of a time‐averaged midposition CT scan for radiotherapy planning of lung cancer patients using deformable registrationa)
  • Aug 11, 2008
  • Medical Physics
  • J W H Wolthaus + 3 more

lower lobe lung tumors move with amplitudes of up to 2 cm due to respiration. To reduce respiration imaging artifacts in planning CT scans, 4D imaging techniques are used. Currently, we use a single (midventilation) frame of the 4D data set for clinical delineation of structures and radiotherapy planning. A single frame, however, often contains artifacts due to breathing irregularities, and is noisier than a conventional CT scan since the exposure per frame is lower. Moreover, the tumor may be displaced from the mean tumor position due to hysteresis. The aim of this work is to develop a framework for the acquisition of a good quality scan representing all scanned anatomy in the mean position by averaging transformed (deformed) CT frames, i.e., canceling out motion. A nonrigid registration method is necessary since motion varies over the lung. 4D and inspiration breath-hold (BH) CT scans were acquired for 13 patients. An iterative multiscale motion estimation technique was applied to the 4D CT scan, similar to optical flow but using image phase (gray-value transitions from bright to dark and vice versa) instead. From the (4D) deformation vector field (DVF) derived, the local mean position in the respiratory cycle was computed and the 4D DVF was modified to deform all structures of the original 4D CT scan to this mean position. A 3D midposition (MidP) CT scan was then obtained by (arithmetic or median) averaging of the deformed 4D CT scan. Image registration accuracy, tumor shape deviation with respect to the BH CT scan, and noise were determined to evaluate the image fidelity of the MidP CT scan and the performance of the technique. Accuracy of the used deformable image registration method was comparable to established automated locally rigid registration and to manual landmark registration (average difference to both methods < 0.5 mm for all directions) for the tumor region. From visual assessment, the registration was good for the clearly visible features (e.g., tumor and diaphragm). The shape of the tumor, with respect to that of the BH CT scan, was better represented by the MidP reconstructions than any of the 4D CT frames (including MidV; reduction of "shape differences" was 66%). The MidP scans contained about one-third the noise of individual 4D CT scan frames. We implemented an accurate method to estimate the motion of structures in a 4D CT scan. Subsequently, a novel method to create a midposition CT scan (time-weighted average of the anatomy) for treatment planning with reduced noise and artifacts was introduced. Tumor shape and position in the MidP CT scan represents that of the BH CT scan better than MidV CT scan and, therefore, was found to be appropriate for treatment planning.

  • Research Article
  • Cite Count Icon 67
  • 10.1118/1.4795133
Modeling respiratory motion for reducing motion artifacts in 4D CT images
  • Mar 20, 2013
  • Medical Physics
  • Yongbin Zhang + 5 more

Four-dimensional computed tomography (4D CT) images have been recently adopted in radiation treatment planning for thoracic and abdominal cancers to explicitly define respiratory motion and anatomy deformation. However, significant image distortions (artifacts) exist in 4D CT images that may affect accurate tumor delineation and the shape representation of normal anatomy. In this study, the authors present a patient-specific respiratory motion model, based on principal component analysis (PCA) of motion vectors obtained from deformable image registration, with the main goal of reducing image artifacts caused by irregular motion during 4D CT acquisition. For a 4D CT image set of a specific patient, the authors calculated displacement vector fields relative to a reference phase, using an in-house deformable image registration method. The authors then used PCA to decompose each of the displacement vector fields into linear combinations of principal motion bases. The authors have demonstrated that the regular respiratory motion of a patient can be accurately represented by a subspace spanned by three principal motion bases and their projections. These projections were parameterized using a spline model to allow the reconstruction of the displacement vector fields at any given phase in a respiratory cycle. Finally, the displacement vector fields were used to deform the reference CT image to synthesize CT images at the selected phase with much reduced image artifacts. The authors evaluated the performance of the in-house deformable image registration method using benchmark datasets consisting of ten 4D CT sets annotated with 300 landmark pairs that were approved by physicians. The initial large discrepancies across the landmark pairs were significantly reduced after deformable registration, and the accuracy was similar to or better than that reported by state-of-the-art methods. The proposed motion model was quantitatively validated on 4D CT images of a phantom and a lung cancer patient by comparing the synthesized images and the original images at different phases. The synthesized images matched well with the original images. The motion model was used to reduce irregular motion artifacts in the 4D CT images of three lung cancer patients. Visual assessment indicated that the proposed approach could reduce severe image artifacts. The shape distortions around the diaphragm and tumor regions were mitigated in the synthesized 4D CT images. The authors have derived a mathematical model to represent the regular respiratory motion from a patient-specific 4D CT set and have demonstrated its application in reducing irregular motion artifacts in 4D CT images. The authors' approach can mitigate shape distortions of anatomy caused by irregular breathing motion during 4D CT acquisition.

  • Research Article
  • 10.3760/cma.j.issn.1004-4221.2013.06.008
A comparative study of GTV displacements of primary thoracic esophageal carcinoma based on repeated 4DCT scans during radiotherapy
  • Nov 15, 2013
  • Chinese Journal of Radiation Oncology
  • Jinzhi Wang + 5 more

Objective To investigate the target volume displacements of primary thoracic esophageal carcinoma (TEC) based on repeated enhanced four-dimensional computed tomography (4DCT) scans during fractionated radiotherapy.Methods Twenty-nine patients with TEC underwent enhanced 4DCT before and in the 10th,20th,and 30th fractions of radiotherapy to delineate the gross tumor volumes (GTVs) and internal gross tumor volumes (IGTVs) in all phases.The GTV displacements of upper,middle,or lower TEC in three-dimensional directions in each time of 4DCT were compared,and the GTV displacements of upper,middle,or lower TEC in the same direction in all 4DCT scans were also compared.The changes in the centroid positions and volumes of IGTV during radiotherapy were determined.Results For the patients with middle TEC,significant differences were found between the GTV displacements in leftright (LR) direction,anterior-posterior (AP) direction,and superor-inferior (SI) direction when 4DCT was performed before or in the 20th fraction of radiotherapy (P =0.000-0.016),and significant differences were found between the GTV displacements in SI direction and LR and AP directions when 4DCT was performed in the 10th fraction of radiotherapy (P =0.000-0.006).For the patients with lower TEC,there were significant differences between the GTV displacements in SI direction and AP direction when 4DCT was performed before or in the 10th or 20th fraction of radiotherapy (P =0.004-0.013).There were no significant differences between the GTV displacements in the same direction in all 4DCT scans (P =0.102-0.823).There were no significant changes in the centroid positions of IGTV during radiotherapy (P =0.6 8 9-0.999),and the most significant decreases in IGTV volumes were seen in the 20th fraction of radiotherapy (P =0.012-0.029).Conclusions Under free breathing,the GTV displacement of upper,middle,or lower TEC in the same direction shows no significant changes in different 4DCT scans during radiotherapy;the IGTV volumes decrease significantly in the 20th fraction of radiotherapy,but there are no significant changes in the centroid positions of IGTV during radiotherapy. Key words: Esophageal neoplasm/ radiotherapy; Tomography, X-ray computed, four-dimensional; Gross tumor volume; Intrafraction target displacement

  • Research Article
  • Cite Count Icon 39
  • 10.1118/1.2739815
The intrafraction motion induced dosimetric impacts in breast 3D radiation treatment: A 4DCT based study
  • Jun 13, 2007
  • Medical Physics
  • Ning J Yue + 5 more

The question remains regarding the dosimetric impact of intrafraction motion in 3D breast treatment. This study was conducted to investigate this issue utilizing the 4DCT scan. The 4D and helical CT scan sets were acquired for 12 breast cancer patients. For each of these patients, based on the helical CT scan, a conventional 3D conformal plan was generated. The breast treatment was then simulated based on the 4DCT scan. In each phase of the 4DCT scan, dose distribution was generated with the same beam parameters as the conventional plan. A software package was developed to compute the cumulative dose distribution from all the phases. Since the intrafraction organ motion is reflected by the 4DCT images, the cumulative dose computed based on the 4DCT images should be closer to what the patient received during treatment. Various dosimetric parameters were obtained from the plan and 4D cumulative dose distribution for the target volume and heart, and were compared to deduce the motion-induced impacts. The studies were performed for both whole breast and partial breast treatment. In the whole breast treatment, the average intrafraction motion induced changes in D95, D90, V100, V95, and V90 of the target volume were -5.4%, -3.1%, -13.4%, -5.1%, and -3.2%, respectively, with the largest values at -26.2%, -14.1%, -91.0%, -15.1%, and -9.0%, respectively. Motion had little impact on the Dmax of the target volume, but its impact on the Dmin of the target volume was significant. For left breast treatment, the motion-induced Dmax change to the heart could be negative or positive, with the largest increase at about 6 Gy. In partial breast treatment, the only non-insignificant impact was in the Dmin of the CTV (ranging from -15.2% to 11.7%). The results showed that the intrafraction motion may compromise target dose coverage in breast treatments and the degree of that compromise was correlated with motion magnitude. However, the dosimetric impact of the motion on the heart dose may be limited.

  • Research Article
  • 10.3760/cma.j.issn.1004-4221.2017.07.015
Simulation of the discrepancy in 4DCT-based cumulative dose using digital phantoms
  • Jul 15, 2017
  • Chinese Journal of Radiation Oncology
  • Gang Liu + 8 more

Objective To conduct a computer simulation to evaluate the discrepancy between the cumulative doses calculated by four-dimensional computed tomography (4DCT) images and 4DCT scans (for real-time respiratory motions) due to the patient’s irregular breathing. Methods A series of digital phantoms were generated from a patient’s 4DCT images to simulate 4DCT images and 4DCT scans (for real-time respiratory motions) resulting from various irregular breathing curves. A six-field intensity-modulated radiotherapy plan was generated. Two cumulative doses in the target were calculated. The first one, named Dall, was calculated by tracking the point displacements in the target manifested on the 4DCT images; the second one, named D4D, was calculated based on the point displacements along the whole breathing motion during 4DCT scanning. Dose discrepancy between D4D and Dall was calculated to evaluate the correlation between breathing pattern and dose discrepancy in the target. Results The dose discrepancy in the target was correlated with mean motion excursion and the standard deviation of motion excursion.ΔDmin(ΔD99) in the target increased from 2.39%(2.04%) to 11.91%(5.24%) as the mean motion excursion increased from 5 mm to 15 mm, and increased from 5.93%(2.15%) to 14.65%(5.01%) as the standard deviation of motion excursion increased from 15% to 45% of the mean motion excursion. When the mean period increased from 3 s to 5 s or the standard deviation of period increased from 10% to 40% of the mean period, ΔDmin(ΔD99) in the target was greater than 6.0%(2.0%), but less than 9.0%(3.0%). When the target diameter was 2 cm, 3 cm, and 4 cm, ΔDmin(ΔD99) in the target was 11.88%(5.50%), 6.91%(2.42%), and 7.53%(3.62%), respectively. Conclusions There is a large discrepancy between the cumulative doses calculated using 4DCT images and 4DCT scans (for real-time respiratory motions) when the patient has irregular breathing. This dose discrepancy depends on mean motion excursion and the standard deviation of motion excursion, but has little relationship with mean period, the standard deviation of period, and tumor volume. Key words: Tomography, X-ray computed; Respiratory-induced motion; Accumulative dose

  • Research Article
  • Cite Count Icon 275
  • 10.1016/j.ijrobp.2008.06.1937
Retrospective Analysis of Artifacts in Four-Dimensional CT Images of 50 Abdominal and Thoracic Radiotherapy Patients
  • Sep 25, 2008
  • International journal of radiation oncology, biology, physics
  • Tokihiro Yamamoto + 4 more

Retrospective Analysis of Artifacts in Four-Dimensional CT Images of 50 Abdominal and Thoracic Radiotherapy Patients

  • Research Article
  • 10.1016/j.jseint.2024.08.182
Magnetic resonance imaging vs. two- and three-dimensional computed tomography scans for assessment of glenoid inclination and version
  • Aug 28, 2024
  • JSES International
  • Thomas Wittmann + 5 more

IntroductionAccurate glenoid component placement is crucial for anatomic (TSA) or reverse (RSA) total shoulder arthroplasty. Preoperative glenoid assessment by using CT scans with or without planning software seems to be the established method to plan implant positions. MRI scans can also display the glenoid bone for preoperative assessment while reducing radiation exposure. Therefore, the objective of this study was to manually assess the glenoid version and inclination in 2D MRI and CT scans in cases with degenerative shoulder pathologies. The results were compared to those of an automated 3D planning software to validate the imaging modality for preoperative glenoid assessment. MethodsMRI and CT scans of 146 patients (n=41 aTSA; n=105 RSA) were included in this retrospective, single-center study. Glenoid version and inclination were measured manually according to Friedman et al and Maurer et al on CT and MRI scans by two observers. Subsequently, the results were compared to the automated measurements performed by a planning software. A repeated-measures analysis of variance (ANOVA) was performed to compare the measured angles and interobserver and intraobserver reliability was calculated using the intraclass correlation coefficients. The level of significance was set p<0.05. ResultsThe average glenoid inclination measured in CT scans was 7.94°±7.33°, in MRI scans 8.56°±7.34° and in automated planning software 7.87°±7.60°. The ANOVA analysis revealed significant differences in mean inclination between 2D MRI and 2D CT (p<0.0005) and between MRI and automated software (p=0.011). No significant difference was found between 2D CT scans and automated planning software (p=1.000). Mean glenoid version measured in 2D CT scans was -7.94°±10.86°, in 2D MRI scans it was -8.04°±10.80° and -8.32°±11.53° by the automated planning software. There was no significant difference in between measurement methods (p = 0.339). Interobserver error analysis showed no statistical differences between the two observers. All measurements had excellent intraobserver reliability. ConclusionPreoperative assessment of glenoid version and inclination is crucial in ensuring precise implant positioning and orientation in TSA and RSA. This study observed a significant level of concordance between manual and automated measuring techniques utilizing MRI and CT scans. Mean glenoid inclination exhibited a statistically significant difference of less than 1° across the assessment modalities and no difference for glenoid version was noted. It seems to be questionable if this finding is clinically relevant. MRI may serve as a viable and safe option for assessing glenoid morphology, version and inclination if CT scans are not available.

  • Research Article
  • Cite Count Icon 39
  • 10.1002/mp.13632
Intelligent 4D CT sequence scanning (i4DCT): Concept and performance evaluation.
  • Jun 23, 2019
  • Medical Physics
  • René Werner + 4 more

Four-dimensional (4D) CT imaging is a central part of current treatment planning workflows in 4D radiotherapy (RT). However, clinical 4D CT image data often suffer from severe artifacts caused by insufficient projection data coverage due to the inability of current commercial 4D CT imaging protocols to adapt to breathing irregularity. We propose an intelligent sequence mode 4D CT imaging protocol (i4DCT) that builds on online breathing curve analysis and respiratory signal-guided selection of beam on/off periods during scan time in order to fulfill projection data coverage requirements. i4DCT performance is evaluated and compared to standard clinical sequence mode 4D CT (seq4DCT) and spiral 4D CT (spiral4DCT) approaches. i4DCT consists of three main blocks: (a) an initial learning period to establish a patient-specific reference breathing cycle representation for data-driven i4DCT parameter selection, (b) online respiratory signal-guided sequence mode scanning (i4DCT core), (c) rapid breathing record analysis and quality control after scanning to trigger potential local rescanning (i4DCT rescan). Based on a phase space representation of the patient's breathing signal, i4DCT core implements real-time analysis of the signal to appropriately switch on and off projection data acquisition even during irregular breathing. Performance evaluation was based on 189 clinical breathing records acquired during spiral 4D CT scanning for RT planning (data acquisition period: 2013-2017; Siemens Somatom with Varian RPM system). For each breathing record, i4DCT, seq4DCT, and spiral4DCT scanning protocol variants were simulated. Evaluation measures were local projection data coverage ; number of local projection data coverage failures; and number of patients with coverage failures; average beam on time as a surrogate for imaging dose and total patient on table time as the time between first and last beam on signal. Using i4DCT, mean inhalation and exhalation projection data coverage increased significantly compared to standard spiral 4D CT scanning as applied for the original clinical data acquisition and conventional 4D CT sequence scanning modes. The improved projection data coverage translated into a reduction of coverage failures by 89% without and 93% when allowing for a rescanning at up to five z-positions compared to spiral scanning and between 76% and 82% without and 85% and 89% with rescanning when compared to seq4DCT. Similar numbers were observed for . Simultaneously, i4DCT (without rescanning) reduced the beam on time on average by 3%-17% compared to standard spiral 4D CT. In turn, the patient on table time increased by between 35% and 66%. Allowing for rescanning led on average to additional 5.9 s beam on and 10.6 s patient on table time. i4DCT outperformed currently implemented clinical fixed beam on period 4D CT scanning approaches by means of a significantly smaller data coverage failure rate without requiring additional beam on time compared to, for example, conventional spiral 4D CT protocols.

  • Research Article
  • 10.3760/cma.j.issn.1004-4221.2014.06.009
Variations of spatial position and overlap ratio for GTV50 and IGTV of primary thoracic esophageal cancer during radiotherapy:a study based on 4DCT scans
  • Nov 15, 2014
  • Chinese Journal of Radiation Oncology
  • 王金之 + 6 more

Objective To investigate the variations of the spatial position and overlap ratio for gross tumor volume (respiratory phase 50%) (GTV50) and internal gross tumor volume (IGTV) of primary thoracic esophageal cancer during conventional fractionated radiotherapy based on repeated four-dimensional computed tomography (4DCT) scans.Methods Thirty-three patients with thoracic esophageal cancer underwent contrast-enhanced 4DCT scans before radiotherapy and at the 10th and 20th fractions of radiotherapy.Scans were registered to the baseline 4DCT scan using bony landmarks.The GTV50 was delineated by the same radiotherapist on each 4DCT imaging data set,and the IGTV was constructed accordingly.The target volume,degree of inclusion (DI),and matching index (MI) were compared in different phases.Results The volumes of GTV50 and IGTV decreased along with treatment course.No significant changes in the centroid position were observed for the GTV50 and IGTV.The median DIs of the target volumes at the 10th and 20th fractions in the original target volume were 0.75 and 0.63(P =0.000) for GTV50 and were 0.79 and 0.66(P=0.000) for IGTV,while the median MIs were 0.61 and 0.56(P=0.002) for GTV50 and were 0.68 and 0.58 (P =0.005) for IGTV.A positive correlation between the variation of volume ratio and the variation of DI was found for GTV50 and IGTV (r =0.632,r =0.783),and the variation of volume ratio was also positively correlated with the variation of MI (r =0.387,r =0.483) ;the 3D vector was negatively correlated with the MI (r =-0.455,r =-0.438).Conclusions During conventional fractionated radiotherapy,the variation of spatial position is less than 0.8 cm for GTV50 and IGTV of primary thoracic esophageal cancer,and the decline of the target leads to varying degrees of decreases in DI and the MI. Key words: Esophageal neoplasm/radiotherapy; Tomography, X-ray computed, four-dimensional; Gross tumor volume; Degree of inclusion; Matching index

  • Research Article
  • 10.3760/cma.j.issn.1004-4221.2014.01.010
Correlation between displacement of GTV and volumetric parameters of primary tumor in thoracic esophageal cancer based on repeated 4DCT scans during radiotherapy
  • Jan 15, 2014
  • Chinese Journal of Radiation Oncology
  • Jinzhi Wang + 7 more

Objective To investigate the correlation between the displacement of gross tumor volume (GTV) and the volume,length,and largest diameter of primary tumor in thoracic esophageal cancer based on repeated enhanced four-dimensional computed tomography (4DCT) scans during fractionated radiotherapy.Methods Thirty enrolled patients with thoracic esophageal cancer underwent enhanced 4DCT scans before radiotherapy and every ten fractions during radiotherapy.The displacements of GTV in left-right (LR),anterior-posterior (AP),and superior-inferior (SI) directions in each scan were obtained,and then the correlation between the displacements and the volume,length,and largest diameter of primary tumor was analyzed.Results In the 20th fraction,a significant positive correlation was observed between the displacement of GTV in LR direction and the volume of primary tumor for all patients and the patients with lower-thoracic esophageal cancer (P =0.012 and 0.040),between the displacement of GTV in SI direction and the length of primary tumor for all patients and the patients with upper-and middle-thoracic esophageal cancer (P =0.003,0.031,and 0.044),and between the displacement of GTV in LR direction and the length of primary tumor for the patients with lower-thoracic esophageal cancer (P =0.027).At the first 4DCT scan before radiotherapy,a significant positive correlation was observed between the largest diameter of primary tumor and the displacement of GTV in LR and SI directions and three-dimensional vector for all patients (P=0.036,0.033,and 0.018) and between the largest diameter of primary tumor and the displacement of GTV in LR direction for the patients with lower-thoracic esophageal cancer (P =0.011).Conclusions In different fractions of radiotherapy,no significant correlation is found between the displacement of GTV in AP direction and the volume,length,and largest diameter of primary tumor in patients with lower-,middle-,or upper-thoracic esophageal cancer.However,their correlation with the displacement of GTV in LR and SI directions depends on the location of tumor and fraction of radiotherapy. Key words: Esophageal neoplasm/radiotherapy; Tomography, X-ray computed, four-dimensional; Gross tumor volume; Correlation of target displacement

  • Abstract
  • Cite Count Icon 2
  • 10.1016/j.ejmp.2016.11.102
50. Magnetic Resonance Imaging optimization for liver SBRT: Breath-triggered acquisition in treatment position to improve lesion contouring
  • Dec 1, 2016
  • Physica Medica
  • S Ken + 7 more

50. Magnetic Resonance Imaging optimization for liver SBRT: Breath-triggered acquisition in treatment position to improve lesion contouring

  • Abstract
  • Cite Count Icon 4
  • 10.1016/j.ijrobp.2008.06.239
Analysis of Artifacts in Four-dimensional CT Images of 50 Abdominal and Thoracic Radiotherapy Patients
  • Aug 20, 2008
  • International Journal of Radiation Oncology*Biology*Physics
  • T Yamamoto + 2 more

Analysis of Artifacts in Four-dimensional CT Images of 50 Abdominal and Thoracic Radiotherapy Patients

  • Research Article
  • Cite Count Icon 12
  • 10.1148/radiol.230193
Assessment of Scapholunate Instability on 4D CT Scans in Patients with Inconclusive Conventional Images.
  • Sep 1, 2023
  • Radiology
  • Sinan Orkut + 7 more

Background Initial imaging work-up using radiography and CT arthrography sometimes can be insufficient to identify a scapholunate (SL) instability (SLI) in patients suspected of having SL ligament tears. Purpose To determine the diagnostic performance of four-dimensional (4D) CT in the identification of SLI and apply the findings to patients suspected of having SLI and with inconclusive findings on radiographs and CT arthrograms. Materials and Methods This prospective single-center study enrolled participants suspected of having SLI (recent trauma, dorsal pain, positive Watson test results, decreased grip strength) between March 2015 and March 2020. Participants with wrist fractures, substantial joint stiffness, or history of wrist surgery were excluded. Each participant underwent radiography, CT arthrography, and 4D CT on the same day. Participants were divided into three groups: those with no SLI, those with SLI, and those with inconclusive results. SL gap and radioscaphoid and lunocapitate angle were measured using semiautomatic quantitative analysis of 4D CT images by two independent readers. Receiver operating characteristic curves were used to evaluate the diagnostic performance of 4D CT. Thresholds were determined with the Youden index and were applied to the inconclusive group. Results Of the 150 included participants (mean age, 41 years ± 14 [SD]; 102 male, 48 female), there were 63 with no SLI, 48 with SLI, and 39 with inconclusive results. The maximum value and range of SL gap measurements on 4D CT scans showed high sensitivity (83% [40 of 48] and 90% [43 of 48], respectively) and high specificity (95% [59 of 62] and 81% [50 of 62], respectively) in the identification of SLI. At least one of these parameters was abnormal on 4D CT scans in 17 of 39 (44%) participants in the inconclusive group, and 10 of 17 (59%) participants had confirmed SLI. In the 22 participants in the inconclusive group with no indication of SLI at 4D CT, follow-up showed no evidence of SLI in 10 (45%) and enabled confirmation of SLI via arthroscopy in three (14%). Conclusion Scapholunate gap measurements on kinematic 4D CT scans enabled correct identification of SLI in 59% of participants with inconclusive results on conventional images. ClinicalTrials.gov registration no. NCT02401568 © RSNA, 2023 Supplemental material is available for this article. See also the editorial by Demehri and Ibad in this issue.

Save Icon
Up Arrow
Open/Close
Notes

Save Important notes in documents

Highlight text to save as a note, or write notes directly

You can also access these Documents in Paperpal, our AI writing tool

Powered by our AI Writing Assistant