Abstract

The study was aimed to conduct a pharmacokinetic evaluation of Raltegravir in rabbit plasma using a sensitive HPLC method. The plasma samples were assayed by Waters alliance e-2695 HPLC instrument using X-bridge phenyl hexyl column (150x4.6mm, 5 µ) under isocratic condition. Acetonitrile and Tri ethyl amine buffer (30:70 v/v) was used as mobile phase with a flow rate of 1 ml/min. The eluent was monitored at 312 nm for measurement of Raltegravir. The calibration curve was linear in the range of 25-1000 ng/ml. The application denotes all the parameters of system suitability, specificity, linearity and accuracy are in good agreement with USFDA guidelines and applied effectively for the investigation of pharmacokinetic studies in rabbit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.