Abstract

We discuss the effects of neutron scattering and γ-ray background in fast neutron imaging and the method to reduce them. As a profiling device, a combination of an imaging plate (IP) and a polypropylene film (CH 2) has been employed in this study. Good profiles were obtained by employing appropriate neutron energy, a CH 2 thickness and geometry for accelerator-based fast neutrons (5–14 MeV). The neutron flux was ∼3.5×10 4 cm −2 s −1 in the CH 2–IP position. Furthermore, we are designing the device using a position-sensitive photomultiplier in order to improve the signal-to-noise ratio by obtaining the information of pulse height for particle selection together with position.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.