Abstract

The RapidHIT™ 200 device from IntegenX® provides a sample-to-profile platform that is capable of processing a variety of sample types. In this study we review the sensitivity of the ‘Run Other’ protocol for processing crime stain type samples containing various input quantities of DNA using the AmpFℓSTR® NGMSElect™ Express PCR Amplification Kit cartridges available from IntegenX®. The range of DNA inputs which achieved useable results were not as desired and therefore various enhancements to the instruments extraction processes were investigated. These studies showed an improvement in the range of DNA input templates that could by processed on the RapidHIT™ 200 by using the enhanced methods and resulted in three new run protocols.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.