Self-powered detectors have been developed with bismuth (2 mm diameter×6.4 mm) and platinum (2 mm diameter ×5 mm) emitters. Tests at a 60Co irradiation facility in 0.125 MR/h (484.5 MC/kg/h) gamma field showed that the gamma sensitivity of the two detectors is 0.121 fA/R/h (0.468 pA/C/kg/h) and 0.17 fA/R/h (0.658 pA/C/kg/h). When tested in the core location in Apsara Swimming Pool reactor the current from Bi detector was assumed to be totally gamma-induced. This information was used to derive the neutron-induced component of the Pt detector in which 55% of the total signal was gamma-induced. The neutron sensitivity was found to be 0.54×10 −21 A/nv.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.