Abstract
A dispersive liquid-liquid microextraction (DLLME) technique based on a solidification-of-floating-organic-drop (SFOD) procedure was developed for the determination of trace amounts of cadmium (Cd) by using a flame atomic absorption spectrometer (FAAS) fitted with a slotted quartz tube (SQT). The extraction of Cd was achieved by forming a complex with diphenylcarbazone. Parameters affecting the formation of complex and extraction outputs were carefully optimized to obtain high-absorbance signals to achieve lower LODs. An SQT was fitted on top of the flame burner head to further enhance the absorbance of the signals recorded by the FAAS. Coupling the DLLME-SFOD procedure with SQT-FAAS produced an enhancement factor of about 183. The LOD of the method was 0.23 µg/L with an RSD of 3.8%. Matrix-matching was used to overcome any low recovery results obtained with tap water and municipal wastewater.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.