Abstract

Objective To develop a small-animal SPECT system using a low cost commercial portable gamma camera equipped with a pinhole collimator, a continuous scintillation crystal and a position-sensitive photomultiplier tube. Material and methods The gamma camera was attached to a variable radius system, which enabled us to optimize sensitivity and resolution by adjusting the radius of rotation to the size of the object. To investigate the capability of the SPECT system for small animal imaging, the dependence of resolution and calibration parameters on radius was assessed and acquisitions of small phantoms and mice were carried out. Results Resolution values, ranging from 1.0 mm for a radius of 21.4 mm and 1.4 mm for a radius of 37.2 mm were obtained, thereby justifying the interest of a variable radius SPECT system. Conclusions The image quality of phantoms and animals were satisfactory, thus confirming the usefulness of the system for small animal SPECT imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.