Abstract

An optimised procedure has been developed for the routine analysis of volatile fatty acids in wastewater matrices, using static headspace gas chromatography with flame ionisation detection. Factors such as sample volume, sample pre-treatment and the time and temperature of sample equilibration have been included in an optimisation model designed to provide maximum detector response for acetic, propionic, iso- and n-butyric and iso- and n-valeric acids in the concentration range 0–1000 mg/l. Optimal headspace conditions were observed when equilibrating at 85 °C for 30 min, using a 2.0 ml sample volume with the addition of 1.0 ml of NaHSO 4 (62%, w/v) into standard 22.3 ml vials. 2-Ethylbutyric acid was used as an internal standard. The suitability of ordinary least squares regression and weighted least squares regression models for the purposes of calibration and quantification were investigated. A weighted least squares linear regression model applied to the heteroscedastic data provided lower detection limits, e.g. 3.7 and 3.3 mg/l for acetic and propionic acids.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.