Abstract

Molybdenum co-factor deficiencies and isolated sulfite oxidase deficiency are rare autosomal recessively inherited diseases characterized by severe psychomotor impairment, intractable seizures, dislocated lens and dysmorphic facial features. The biochemical diagnosis of these diseases requires the determination of urine sulfocysteine. Urine sulfocysteine was quantified by an ultra-high performance liquid chromatography-MS/MS assay. The method was validated for linearity, accuracy, precision, recovery and stability. Total imprecision of accuracy was less than 6%. Intra-assay and inter-assay precisions were less than 5%. The recovery was higher than 98%. The method is inexpensive, fast, accurate and has been successfully used for identifying five molybdenum co-factor deficient and six sulfite oxidase deficient patients since deployed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.