Abstract

Peripheral subtraction scanning is used to trace the blood vessels of upper and lower extremities. In some modern C-arm fluoroscopy systems this function is performed automatically. In this mode the system is programmed to advance and stop in a series of steps taking a mask image at each point. The system then repeats each step after the contrast agent has been injected, and produces a DSA image at each point. Current radiographic quality assurance protocols do not address this feature. This note reviews methods of measuring system vibration while images are being acquired in automated peripheral stepping. The effect on image quality pre- and post-image processing is assessed. Results show that peripheral stepping DSA does not provide the same degree of image quality as static DSA. In examining static test objects, the major cause of the reduction in image quality is misregistration due to vibration of the image intensifier during imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.