Abstract

The next generation of flexible endoscopy platforms such as The Master and Slave Transluminal Endoscopic Robot (MASTER) is primarily for remote control manipulation of forceps, but manipulation of the flexible endoscope itself still depends on conventional techniques. We have developed the Endoscopic Operation Robot (EOR) ver.3, which incorporates haptic feedback to provide complete remote control flexible-endoscope manipulation. The present study aimed to evaluate the performance of the EOR ver.3. A colonoscopy training model was used with scope insertion to the cecum. Force during insertion and insertion time (seconds) to the cecum were evaluated. The data were compared by colon zone and experience level (trainee or expert). The mean insertion time into the cecum was 118.54 ± 89.42 seconds. Stronger force and torque were required for deeper insertion of the scope. Expert and trainee endoscopists differed in the insertion time to the cecum, maximum counterclockwise torque, mean clockwise torque, and mean counterclockwise torque. The EOR ver.3 has operability with which endoscopists can easily familiarize themselves.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.