We have developed a reversed phase high performance liquid chromatography pulsed amperometric detection (RPHPLC-PAD) method for the determination of paromomycin. It is sensitive, repeatable, and selective without the pretreatment step.Trifluoroacetic acid–water was utilized as the eluent and detected by PAD under NaOH alkaline conditions. The paromomycin detection limit (S/N=3.3) was 2μgmL−1 and the quantification limit (S/N=10) was 6μgmL−1. Coefficients of linear regression were higher than 0.99 for concentrations between 6.25 and 200μgmL−1. The intra and inter-day precision (RSD) was less than 6.5%. The average recoveries were 97.53–102.01%. The proposed HPLC-PAD method presented advantageous performance characteristics and it can be considered suitable for the evaluation of paromomycin loaded nanogel formulation in ex vivo permeation and in vitro release studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.